S2VNA and S4VNA SCPI Programming Manual
273
SENS:CORR:COLL:THRU:ADD:LENG
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:LENGth
<numeric>
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:LENGth?
Description
Sets or reads out the approximate value of the mechanical length
of an unknown thru in the thru addition function. This value is
used to eliminate the uncertainty of ± 180 ° when calculating the
phase response of the thru.
If this value is set to zero, the analyzer uses an algorithm to
automatically determine the delay of the thru. In most cases
setting this value to zero is enough. Setting this value to non zero
is required when:


(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the approximate value of the thru length .
Unit
Meters
Query Response
<numeric>
Preset Value
0
Note
The delay and the length of the thru can be set mutually.



Equivalent Softkeys
Calibration > Calibrate > Thru Addition > Delay