S2VNA and S4VNA SCPI Programming Manual
152
CALC:TRAN:TIME:IMP:WIDT
CALCulate
<Ch>
[:SELected]:TRANsform:TIME:IMPulse:WIDTh
<time>
CALCulate
<Ch>
[:SELected]:TRANsform:TIME:IMPulse:WIDTh?
Description
Sets or reads out the impulse width (time domain transformation
resolution), coupled with the Kaiser–Bessel window shape β
parameter. The impulse width setting changes the β parameter,
and setting of β parameter changes the impulse width.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<time> the impulse width, the range varies depending on the specified
frequency range and the number of points
Unit
s (second)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Equivalent Softkeys
Analysis > Time Domain > Window > Impulse Width (when the
transformation type is set to Bandpass or Lowpass Impulse)