S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:METH:THRU
SENSe
<Ch>
:CORRection:COLLect:METHod[:RESPonse]:THRU
<rcvport>
,
<srcport>
Selects the ports and sets the response calibration (Thru) type for
the calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<rcvport> the number of the receiver port from 1 to 2 (or 4)
<srcport> the number of the source port from 1 to 2 (or 4)
If the same port numbers are specified, an error occurs.
Calibration > Calibrate > Response (Thru) > Select Port
SENS:CORR:COLL:METH:TRL2
SENSe
<Ch>
:CORRection:COLLect:METHod:TRL2
<port1>
,
<port2>
Selects the ports and sets the 2–port TRL calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 2 (or 4)
<port2> the second port number from 1 to 2 (or 4)
If the same port numbers are specified, an error occurs.
Calibration > Calibrate > 2–Port TRL Cal