S2VNA and S4VNA SCPI Programming Manual
267
SENS:CORR:COLL:METH:THRU
SENSe
<Ch>
:CORRection:COLLect:METHod[:RESPonse]:THRU
<rcvport>
,
<srcport>
Description
Selects the ports and sets the response calibration (Thru) type for
the calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<rcvport> the number of the receiver port from 1 to 2 (or 4)
<srcport> the number of the source port from 1 to 2 (or 4)
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > Response (Thru) > Select Port
SENS:CORR:COLL:METH:TRL2
SENSe
<Ch>
:CORRection:COLLect:METHod:TRL2
<port1>
,
<port2>
Description
Selects the ports and sets the 2–port TRL calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 2 (or 4)
<port2> the second port number from 1 to 2 (or 4)
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > 2–Port TRL Cal

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish