S2VNA and S4VNA SCPI Programming Manual
153
CALC:TRAN:TIME:KBES
CALCulate
<Ch>
[:SELected]:TRANsform:TIME:KBESsel
<numeric>
CALCulate
<Ch>
[:SELected]:TRANsform:TIME:KBESsel?
Description
Sets or reads out the β parameter, which controls the Kaiser–
Bessel window shape, when performing time domain
transformation.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> β parameter from 0 to 13
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
6
Equivalent Softkeys
Analysis > Time Domain > Window > Kaiser Beta
CALC:TRAN:TIME:LPFR
CALCulate
<Ch>
[:SELected]:TRANsform:TIME:LPFRequency
Description
Changes the frequency range to match with the low-pass type of
the time domain transformation function.
(no query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
Analysis > Time Domain > Set Frequency Low Pass

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish