S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:METH:OPEN
SENSe
<Ch>
:CORRection:COLLect:METHod[:RESPonse]:OPEN
<port>
Selects the port and sets the response calibration (Open) type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port> the number of the port from 1 to 2 (or 4)
Calibration > Calibrate > Response (Open) > Select Port
SENS:CORR:COLL:METH: SHOR
SENSe
<Ch>
:CORRection:COLLect:METHod[:RESPonse]:SHORt
<port>
Selects the port and sets the response calibration (Short) type for
the calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port> the number of the port from 1 to 2 (or 4)
Calibration > Calibrate > Response (Short) > Select Port