S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:THRU:ADD:FULL3:PORT
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL3:PORTs
<port1>
,
<port2>
,
<port3>
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL3:PORTs?
Selects the pors to comlete the 3–port calibration in the thru
addition function.
(command/query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
SENS:CORR:COLL:THRU:ADD:FULL3:COMP
Calibration > Calibrate > Thru Addition > Comlete 3-Port Calibration >
Select Ports
SENS:CORR:COLL:THRU:ADD:FULL3:ACQ
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL3:ACQuire
<port1>,<port2>
Measures an arbitrary thru between the specified ports. The
measurements are used to complete the 3-port calibration in the
thru addition function by the command
SENS:CORR:COLL:THRU:ADD:FULL3:COMP.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
Calibration > Calibrate > Thru Addition > Comlete 3-Port Calibration >
Thru m-n