Page #206 background imageLoading...
Page #206 background image
S2VNA and S4VNA SCPI Programming Manual
206
SENS:AVER
SENSe
<Ch>
:AVERage[:STATe] {OFF|ON|0|1}
SENSe
<Ch>
:AVERage[:STATe]?
Description
Turns ON/OFF the measurement averaging function.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Related Commands
SENS:AVER:COUN
Equivalent Softkeys
Average > Averaging
SENS:AVER:CLE
SENSe
<Ch>
:AVERage:CLEar
Description
Restarts the averaging process, when the averaging function
turned on.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Related Commands
SENS:AVER
Equivalent Softkeys
None