EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sour:pow:port

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
355
SOUR:POW:PORT
SOURce
<Ch>
:POWer:PORT
<Pt>
[:LEVel][:IMMediate][:AMPLitude]
<power>
SOURce
<Ch>
:POWer:PORT
<Pt>
[:LEVel][:IMMediate][:AMPLitude]?
Description
Sets or reads out the power level of each port for the frequency
sweep type when the port couple feature is set to OFF by the
SOUR:POW:PORT:COUP command.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<power> the power level within the power limits of the analyzer.
Unit
dBm (decibels above 1 milliwatt)
Resolution
0.05 dBm
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
Stimulus > Power > Port Power > Port n

Table of Contents