EasyManuals Logo

Copper Mountain Technologies S2 Series User Manual

Copper Mountain Technologies S2 Series
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #91 background imageLoading...
Page #91 background image
S2VNA and S4VNA SCPI Programming Manual
91
CALC:FILT:TIME:STOP
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:STOP
<time>
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:STOP?
Description
Sets or reads out the gate stop value of the gating function.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<time> the stop value of the gate, the range varies depending on the
frequency span and the number of points
Unit
s (second)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
+1e–8
Equivalent Softkeys
Analysis > Gating > Stop

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish