S2VNA and S4VNA SCPI Programming Manual
304
SENS:CORR:REC:OFFS:AMPL
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:OFFSET:AMPLitude
<numeric>
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:OFFSET:AMPLitude
?
Description
Sets or reads out the power offset value when the Receiver
Calibration is performed. Receiver calibration is done at the
condition of <source power> + < power offset>.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<numeric> the power offset value when the Receiver Calibration is
performed from -100 to 100.
Unit
dBm
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0.0
Equivalent Softkeys
Calibration > Receiver Calibration > Power Offset