EasyManuals Logo

Copper Mountain Technologies S2 Series User Manual

Copper Mountain Technologies S2 Series
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #270 background imageLoading...
Page #270 background image
S2VNA and S4VNA SCPI Programming Manual
270
SENS:CORR:COLL:SAVE
SENSe
<Ch>
:CORRection:COLLect:SAVE
Description
Calculates the calibration coefficients from the calibration
standards measurements depending on the selected calibration
type. The calibration type is selected by one of commands
SENS:CORR:COLL:METH:XXXX.
On completion of the command, all the calibration standards
measurements are cleared and the error correction automatically
turns ON.
At the attempt to execute this command before all the needed
standards are measured, an error occurs and the command is
ignored.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Related Commands
Calibration type selection:
Calibration standards measurement:
SENS:CORR:COLL:ISOL
SENS:CORR:COLL:LOAD
SENS:CORR:COLL:OPEN
SENS:CORR:COLL:SHOR
SENS:CORR:COLL:THRU
SENS:CORR:COLL:TRLT
SENS:CORR:COLL:TRLL
SENS:CORR:COLL:TRLR
Equivalent Softkeys
Calibration > Calibrate > … > Apply

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish