CALC:FILT:TIME:STAR
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:STARt
<time>
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:STARt?
Description
Sets or reads out the gate start value of the gating function.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<time> the start value of the gate, the range varies depending on the
frequency span and the number of points
Unit
s (second)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
1e8
Equivalent Softkeys
Analysis > Gating > Start