EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Calc:filt:time:cent

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
86
CALC:FILT:TIME:CENT
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:CENTer
<time>
CALCulate
<Ch>
[:SELected]:FILTer[:GATE]:TIME:CENTer?
Description
Sets or reads out the gate center value of the gating function.
(command/query)
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<time> the center value of the gate, the range varies depending on the
frequency span and the number of points
Unit
s (second)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
Analysis > Gating > Center

Table of Contents