S2VNA and S4VNA SCPI Programming Manual
299
SENS:CORR:OFFS:COLL:SHOR
SENSe
<Ch>
:CORRection:OFFSet:COLLect[:ACQuire]:SHORt
<numeric1>
,
<numeric2>
Description
Measures the calibration data of the short standard of the specified
port when the frequency offset feature is on for scalar mixer
calibration.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric1> Measurement port number
<numeric2> Frequency port number.
Notes
The command starts the measurement immediately if the trigger source
for calibration set to the "Internal" by the command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration >
Reflection Port n > Port n @Freq m Short