S2VNA and S4VNA SCPI Programming Manual
306
SENS:CORR:TRAN:TIME:LOSS
SENSe
<Ch>
:CORRection:TRANsform:TIME:LOSS
<numeric>
SENSe
<Ch>
:CORRection:TRANsform:TIME:LOSS?
Description
Sets or reads out the cable loss value for the cable correction function,
when the time domain transformation function is turned ON.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the cable loss value.
Unit
dB/m (decibell / meter)
Query Response
<numeric>
Preset Value
0 dB/m
Equivalent Softkeys
Analysis > Time Domain > Cable Correction > Cable Loss
SENS:CORR:TRAN:TIME:RVEL
SENSe
<Ch>
:CORRection:TRANsform:TIME:RVELocity
<numeric>
SENSe
<Ch>
:CORRection:TRANsform:TIME:RVELocity?
Description
Sets or reads out the cable relative wave speed velocity for the cable
correction function, when the time domain transformation function is
turned ON.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the cable velocity factor.
Query Response
<numeric>
Preset Value
1.0
Equivalent Softkeys
Analysis > Time Domain > Cable Correction > Velocity Factor

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish