EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sens:band; Sens:bwid; Sens:corr:cle

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
208
SENS:BAND
SENSe
<Ch>
:BANDwidth[:RESolution]
<frequency>
SENSe
<Ch>
:BANDwidth[:RESolution]?
SENS:BWID
SENSe
<Ch>
:BWIDth[:RESolution]
<frequency>
SENSe
<Ch>
:BWIDth[:RESolution]?
Description
Sets or reads out the IF bandwidth.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<frequency> the IF bandwidth value
Unit
Hz (Hertz)
Out of Range
Sets the value of the analyzer limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
10 kHz
Resolution
In steps of 1, 1.5, 2, 3, 5, 7
Equivalent Softkeys
Average > IF Bandwidth
SENS:CORR:CLE
SENSe
<Ch>
:CORRection:CLEar
Description
Clears the calibration coefficient table.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
None