S2VNA and S4VNA SCPI Programming Manual
354
SOUR:POW:CENT
SOURce
<Ch>
:POWer:CENTer
<power>
SOURce
<Ch>
:POWer:CENTer?
Description
Sets or reads out the center value of the power sweep type.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<power> the power level within the power limits of the analyzer.
Unit
dBm (decibels above 1 milliwatt)
Resolution
0.05 dBm
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
Depends on the analyzer
Equivalent Softkeys
Stimulus > Center