EasyManuals Logo

Copper Mountain Technologies S2 Series User Manual

Copper Mountain Technologies S2 Series
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #385 background imageLoading...
Page #385 background image
S2VNA and S4VNA SCPI Programming Manual
385
SYST:CYCL:TIME:MEAS?
SYSTem:CYCLe:TIME:MEASurement?
Description
Reads out the measured cycle time. The cycle time is the interval
between the start of two adjacent sweeps. The cycle time is
averaged by an exponential window with a time constant of about
0.5 sec. If the cycle time is changed more than 100 usec in
comparison with the averaged time, the averaging starts anew.
(query only)
Target
Analyzer
Unit
second
Query Response
<numeric>
Equivalent Softkeys
Display > Properies > Cycle Time
SYST:DATE
SYSTem:DATE
<numeric 1>
,
<numeric 2>
,
<numeric 3>
SYSTem:DATE?
Description
Sets or reads out the current date.
(command/query)
Parameter
<numeric 1> year from 1900 to 2100;
<numeric 2> month from 1 to 12;
<numeric 3> day from 1 to 31.
Query Response
<numeric 1>, <numeric 2>, <numeric 3>
Equivalent Softkeys
None