S2VNA and S4VNA SCPI Programming Manual
216
SENS:CORR:COLL:ADAP:LENG
SENSe
<Ch>
:CORRection:COLLect:ADAPter:LENGth
<numeric>
SENSe
<Ch>
:CORRection:COLLect:ADAPter:LENGth?
Description
Sets or reads out the approximate value of the mechanical length
of the adapter in the adapter removal/insertion function. This value
is used to eliminate the uncertainty of ± 180 ° when calculating
the phase response of the adapter.
The sign of the value depends on the type of the removal /
insertion function. The value must be negative for the adapter
removal function and must be positive for the adapter insertion
function.
If this value is set to zero, the analyzer uses an algorithm to
automatically determine the delay of the adapter. In most cases
setting this value to zero is enough. Setting this value to non zero
is required when:


(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the approximate length of the adapter (with minus sign
when adapter is removed).
Unit
Meters
Query Response
<numeric>
Preset Value
0
Note
The delay and the length of the adapter can be set mutually.



Equivalent Softkeys
Calibration > Calibrate > Adapter Removal / Insertion > Adapter Delay