S2VNA and S4VNA SCPI Programming Manual
283
SENS:CORR:EXT:AUTO:DCOF
SENSe
<Ch>
:CORRection:EXTension:AUTO:DCOFfset {OFF|ON|0|1}
SENSe
<Ch>
:CORRection:EXTension:AUTO:DCOFfset?
Description
Turns ON/OFF the usage of "Loss at DC" value for the results of the
auto port extension function.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Port Extension > Auto Port Extension > Adjust Mismatch
{OFF | ON }
SENS:CORR:EXT:AUTO:LOSS
SENSe
<Ch>
:CORRection:EXTension:AUTO:LOSS {OFF|ON|0|1}
SENSe
<Ch>
:CORRection:EXTension:AUTO:LOSS?
Description
Turns ON/OFF the usage of "Loss1"and "Loss2" values for the
results of the auto port extension function.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Port Extension > Auto Port Extension > Adjust Mismatch
{OFF | ON }