Loading...
S2VNA and S4VNA SCPI Programming Manual
210
SENS:CORR:COEF:METH:ERES
SENSe
<Ch>
:CORRection:COEFficient:METHod:ERESponse
<rcvport>
,
<srcport>
Description
Selects the ports and sets the 1path 2port calibration type, when
the written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<rcvport> the number of the receiver port from 1 to 2 (or 4)
<srcport> the number of the source port from 1 to 2 (or 4)
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COEF:SAVE
Equivalent Softkeys
None
SENS:CORR:COEF:METH:OPEN
SENSe
<Ch>
:CORRection:COEFficient:METHod[:RESPonse]:OPEN
<port>
Description
Selects the port and sets the response calibration (Open) type, when
the written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port> the number of the port from 1 to 2 (or 4)
Related Commands
SENS:CORR:COEF:SAVE
Equivalent Softkeys
None

Table of Contents