S2VNA and S4VNA SCPI Programming Manual
242
SENS:CORR:COLL:CKIT:STAN:TYPE
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:TYPE
<char>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:TYPE?
Description
Sets or reads out the type of calibration standard.
(command/query)
Target
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N the number of the standards in
the calibration kit
Parameter
<char> Specifies the type of calibration standard:
OPEN : Open
SHORt : Short
LOAD : Load
THRU : Thru
NONE : Not defined
Query Response
{OPEN|SHOR|LOAD|THRU|NONE}
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > STD Type