S2VNA and S4VNA SCPI Programming Manual
201
MMEM:STOR:SNP
MMEMory:STORe:SNP[:DATA]
<string>
Description
Saves the measured S–parameters of the active channel into a
Touchstone file. The file type (1–port to 4–port) is set by the
MMEM:STOR:SNP:TYPE:SxP command.
(no query)
Target
Active channel, set by the DISP:WIND:ACT command
Parameter
<string> File name
Notes
If the full path of the file is not specified, the \FixtureSim subdirectory of
the application directory will be searched for the file. The file has *.sNp
extension by default.
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Save File…
MMEM:STOR:SNP:FORM
MMEMory:STORe:SNP:FORMat
<char>
MMEMory:STORe:SNP:FORMat?
Description
Sets the data format for the S–parameter saving by the
MMEM:STOR:SNP command.
(command/query)
Parameter
<char> Choose from:
MA : Logarithmic Magnitude / Angle format
DB : Linear Magnitude / Angle format
RI : Real part /Imaginary part format
Query Response
{RI|DB|MA}
Preset Value
RI
Equivalent Softkeys
Save/Recall > Save Data to Touchstone File > Format

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Copper Mountain Technologies S2 Series and is the answer not in the manual?

Copper Mountain Technologies S2 Series Specifications

General IconGeneral
BrandCopper Mountain Technologies
ModelS2 Series
CategoryMeasuring Instruments
LanguageEnglish