S2VNA and S4VNA SCPI Programming Manual
MMEM:STOR:SNP
MMEMory:STORe:SNP[:DATA]
<string>
Saves the measured S–parameters of the active channel into a
Touchstone file. The file type (1–port to 4–port) is set by the
MMEM:STOR:SNP:TYPE:SxP command.
(no query)
Active channel, set by the DISP:WIND:ACT command
If the full path of the file is not specified, the \FixtureSim subdirectory of
the application directory will be searched for the file. The file has *.sNp
extension by default.
Save/Recall > Save Data to Touchstone File > Save File…
MMEM:STOR:SNP:FORM
MMEMory:STORe:SNP:FORMat
<char>
MMEMory:STORe:SNP:FORMat?
Sets the data format for the S–parameter saving by the
MMEM:STOR:SNP command.
(command/query)
<char> Choose from:
MA : Logarithmic Magnitude / Angle format
DB : Linear Magnitude / Angle format
RI : Real part /Imaginary part format
Save/Recall > Save Data to Touchstone File > Format