S2VNA and S4VNA SCPI Programming Manual
398
TRIG:AVER
TRIGger[:SEQuence]:AVERage {OFF|ON|0|1}
TRIGger[:SEQuence]:AVERage?
Description
Turns ON/OFF the averaging trigger function. The function
executes a sweep the number of times specified by the averaging
factor with a single trigger for the channels with the averaging
enabled.
The averaging process begins again with each trigger.
Note: The point trigger function has priority against this command.
When the point trigger is enabled the number of pulses equal to
(number of points) x (averaging factor) is needed to complete the
averaging.
(command/query)
Parameter
Specifies the averaging trigger function state:
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Related Commands
SENS:AVER
Equivalent Softkeys
Average > Avg Trigger { ON | OFF }