S2VNA and S4VNA SCPI Programming Manual
408
TRIG:STAT?
TRIGger[:SEQuence]:STATus?
Description
Reads out the the current state of the analyzer.
(query only)
Query Response
HOLD : Hold
MEAS : Measure (sweep in progress)
WAIT : Waiting for trigger
Equivalent Softkeys
None
TRIG:WAIT
TRIGger[:SEQuence]:WAIT
<char>
Description
Delays the execution by the analyzer of the next command till the
specified state of the analyzer has been reached (see options
below).
A query that follows the TRIG:WAIT command blocks the execution
of the user program till the specified state of the analyzer has been
reached.
The command can be used to wait for the end of the sweep
initianed by the commands TRIG, *TRG or initiated by the external
trigger signal. If the continuous initiation mode is turned ON by the
command INIT:CONT ON, then the parameter of the command
must be WAIT, otherwise HOLD.
(no query)
Parameter
<char> Choose from:
HOLD : Waits for the Hold state
MEASure : Waits for the Measure state
WAIT : Waits for the Trigger Waiting state
Related Commands
TRIG, *TRG
TRIG:SOUR EXT
Equivalent Softkeys
None