Delays the execution by the analyzer of the next command till the
specified state of the analyzer has been reached (see options
below).
A query that follows the TRIG:WAIT command blocks the execution
of the user program till the specified state of the analyzer has been
reached.
The command can be used to wait for the end of the sweep
initianed by the commands TRIG, *TRG or initiated by the external
trigger signal. If the continuous initiation mode is turned ON by the
command INIT:CONT ON, then the parameter of the command
must be WAIT, otherwise HOLD.
(no query)
<char> Choose from:
HOLD : Waits for the Hold state
MEASure : Waits for the Measure state
WAIT : Waits for the Trigger Waiting state