EasyManuals Logo

Copper Mountain Technologies S2 Series User Manual

Copper Mountain Technologies S2 Series
428 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #188 background imageLoading...
Page #188 background image
S2VNA and S4VNA SCPI Programming Manual
188
INIT:CONT
INITiate
<Ch>
:CONTinuous {OFF|ON|0|1}
INITiate
<Ch>
:CONTinuous?
Description
Turns ON/OFF the continuous trigger initiation mode.
When the continuous initiation mode turned ON:
ï‚· If the Internal trigger source is selected by the command
TRIG:SOUR INT, then the channel continuously sweeps;
ï‚· If the trigger source other than the internal is selected, then
the channel goes to the trigger waiting state. Upon receipt of
a trigger from the selected source, the sweep starts for the
channels awaiting trigger. On completion of the sweep the
channel goes to the trigger waiting state.
When the continuous trigger initiation mode turned OFF the channel
is in the Hold state, to initiate a sweep one should use the INIT
command.
(command /query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
Specifies the continuous trigger initiation mode:
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
1
Related Commands
TRIG:SOUR
INIT
Equivalent Softkeys
Stimulus > Trigger > Continuous
Stimulus > Trigger > Hold

Table of Contents