Calculates the calibration coefficients for the simplified 3 or 4 port
calibration from the calibration standards measurements when the
3 or 4 port calibration is selected as the calibration type. The
calibration type is selected by one of commands
SENS:CORR:COLL:METH:SOLT3/SOLT4/TRL3/TRL4.
The simplified 3 port calibration allows omit one THRU
measurement. The simplified 4 port calibration allows omit up to
three THRU measurements.
If full set of calibration standard measurement is made this
command behaves like the SENS:CORR:COLL:SAVE command.
(no query)
Calibration type selection:
SENS:CORR:COLL:METH:SOLT3
SENS:CORR:COLL:METH:SOLT4
SENS:CORR:COLL:METH:TRL3
SENS:CORR:COLL:METH:TRL4
Calibration standards measurement:
SENS:CORR:COLL:ISOL
SENS:CORR:COLL:LOAD
SENS:CORR:COLL:OPEN
SENS:CORR:COLL:SHOR
SENS:CORR:COLL:THRU
SENS:CORR:COLL:TRLT
SENS:CORR:COLL:TRLL
SENS:CORR:COLL:TRLR