S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:LOAD
SENSe
<Ch>
:CORRection:COLLect[:ACQuire]:LOAD
<port>
Measures the calibration data of the load standard for the specified
port.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port> the number of the port from 1 to 2 (or 4)
The command starts the measurement immediately if the trigger source
for calibration set to the "Internal" by the command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Calibration > Calibrate > Response (Open) > Load (Optional)
Calibration > Calibrate > Response (Short) > Load (Optional)
Calibration > Calibrate > Full 1–Port Cal > Load
Calibration > Calibrate > One Path 2–Port Cal > Load
Calibration > Calibrate > Full 2–Port Cal > Port n Load