S2VNA and S4VNA SCPI Programming Manual
321
SENS:OFFS:PORT:OFFS
SENSe
<Ch>
:OFFSet:PORT
<Pt>
[:FREQuency]:OFFSet
<frequency>
SENSe
<Ch>
:OFFSet:PORT
<Pt>
[:FREQuency]:OFFSet?
Description
Sets or reads out the basic frequency range offset of port <Pt>
when the frequency offset feature is ON and offset type is "PORT".
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<frequency> offset from 1e12 to 1e12;
Unit
Hz
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Related Commands
SENS:OFFS:STAT
SENS:OFFS:TYPE
Equivalent Softkeys
Stimulus > Frequency Offset > Port n > Offset