SENS:CORR:COLL:OPEN
SENSe
<Ch>
:CORRection:COLLect[:ACQuire]:OPEN
<port>
Description
Measures the calibration data of the open standard for the
specified port.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port> the number of the port from 1 to 2 (or 4)
Notes
The command starts the measurement immediately if the trigger source
for calibration set to the "Internal" by the command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Equivalent Softkeys
Calibration > Calibrate > Response (Open) > Open
Calibration > Calibrate > Full 1Port Cal > Open
Calibration > Calibrate > One Path 2Port Cal > Open
Calibration > Calibrate > Full 2Port Cal > Port n Open