S2VNA and S4VNA SCPI Programming Manual
231
SENS:CORR:COLL:CKIT:STAN:C1
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:C1
<numeric>
SENSe:CORRection:COLLect:CKIT:STAN
<Std>
:C1?
Description
Sets or reads out the C1 value for the open calibration standard.
(command/query)
Target
Standard <Std> of the selected calibration kit,
<Std>={[1]|2|...N}, where N – the number of the standards in
the calibration kit
Parameter
<numeric> the C1 value from –1E18 to 1E18
Unit
1E–27 F/Hz (Farad/Hertz)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Equivalent Softkeys
Calibration > Cal Kit > Define STDs > C1 10
–27
F/Hz