S2VNA and S4VNA SCPI Programming Manual
127
CALC:MARK:FUNC:PPOL
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
:FUNCtion:PPOLarity
<char>
CALCulate
<Ch>
[:SELected]:MARKer
<Mk>
:FUNCtion:PPOLarity?
Description
Selects the peak polarity, when the marker search for peak is
performed by the CALC:MARK:FUNC:EXEC command.
(command/query)
Target
Marker <Mk> of the active trace of channel <Ch>,
<Ch>={[1]|2|...16}
<Mk>={[1]|2|...16}
Parameter
<char> Specifies the peak polarity:
POSitive : Positive polarity
NEGative : Negative polarity
BOTH : Both positive polarity and negative polarity
Query Response
{POS|NEG|BOTH}
Preset Value
POS
Related Commands
CALC:MARK:FUNC:EXEC
Equivalent Softkeys
Markers > Marker Search > Peak > Peak Polarity > { Positive | Negative |
Both }