S2VNA and S4VNA SCPI Programming Manual
227
SENS:CORR:COLL:CKIT:ORD:TRLT
SENSe:CORRection:COLLect:CKIT:ORDer:TRLThru
<port1>
,
<port2>
,
<numeric>
SENSe:CORRection:COLLect:CKIT:ORDer:TRLThru?
<port1>
,
<port2>
Description
Sets or reads out the number of the calibration standard of the TRL
thru type, used for the measurement between the <port1> and
<port2> ports.
(command/query)
Target
Selected calibration kit
Parameter
<port1> the number of the receiver port from 1 to 2 (or 4)
<port2> the number of the source port from 1 to 2 (or 4)
<numeric> the number of the calibration standard
Out of Range
If the specified standard number is greater than the number of standards
in the kit, an error occurs. If the specified standard number is not the thru
standard number, an error occurs.
Equivalent Softkeys
Calibration > Cal Kit > Specify CLSs > TRL Thru Port x-y (Row)