S2VNA and S4VNA SCPI Programming Manual
298
SENS:CORR:OFFS:COLL:PMETer
SENSe
<Ch>
:CORRection:OFFSet:COLLect[:ACQuire]:PMETer
<numeric1>
,
<numeric2>
,
<numeric3>
Description
Measures the scalar-mixer calibration data using the power meter
when the frequency offset feature is ON.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric1> Measurement port number
<numeric2> Frequency port number.
<numeric3> Always 0 (reserved).
Notes
The command starts the measurement of the calibration data
immediately regardless the trigger settings. The command blocks the
execution of the subsequent commands until the completion of the
measurement.
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration >
Power > Port n @Freq m