S2VNA and S4VNA SCPI Programming Manual
266
SENS:CORR:COLL:METH:SOLT3
SENSe
<Ch>
:CORRection:COLLect:METHod:SOLT3
<port1>
,
<port2>
,
<port3>
Description
Selects the ports and sets the full 3port calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > Full 3Port Cal
SENS:CORR:COLL:METH:SOLT4
SENSe
<Ch>
:CORRection:COLLect:METHod:SOLT4
<port1>
,
<port2>
,
<port3>
,
<port4>
Description
Selects the ports and sets the full 4port calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
<port4> the fourth port number from 1 to 4
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > Full 4Port Cal