S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:METH:SOLT3
SENSe
<Ch>
:CORRection:COLLect:METHod:SOLT3
<port1>
,
<port2>
,
<port3>
Selects the ports and sets the full 3–port calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
If the same port numbers are specified, an error occurs.
Calibration > Calibrate > Full 3–Port Cal
SENS:CORR:COLL:METH:SOLT4
SENSe
<Ch>
:CORRection:COLLect:METHod:SOLT4
<port1>
,
<port2>
,
<port3>
,
<port4>
Selects the ports and sets the full 4–port calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
<port4> the fourth port number from 1 to 4
If the same port numbers are specified, an error occurs.
Calibration > Calibrate > Full 4–Port Cal