S2VNA and S4VNA SCPI Programming Manual
CALC:MARK:ACT
CALCulate
<Ch>
:SELected]:MARKer
<Mk>
:ACTivate
Sets the active marker.
If the marker is not ON, this function will turn the marker ON.
Turning ON a marker with the number from 1 to 15 will turn ON all
the markers of smaller numbers. Turning ON the reference marker
with number 16 does not turn ON the markers with the numbers
from 1 to 15, but switchs these markers to the relative
measurement mode.
(no query)
Marker <Mk> of the active trace of channel <Ch>,
<Ch>={[1]|2|...16}
<Mk>={[1]|2|...16}
Markers > Select > Marker n
Markers > Reference Marker
CALC:MARK:BWID
CALCulate
<Ch>
[:SELected]:MARKer:BWIDth[:STATe] {OFF|ON|0|1}
CALCulate
<Ch>
[:SELected]:MARKer:BWIDth[:STATe]?
Turns ON/OFF the bandwidth search function.
(command/query)
The active trace of channel <Ch>,
<Ch>={[1]|2|...16}
{ON|1} : ON
{OFF|0} : OFF
Markers > Marker Math > Bandwidth Search > Bandwidth Search