S2VNA and S4VNA SCPI Programming Manual
281
SENS:CORR:COLL:THRU:ADD:FULL4:COMP
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL4:COMPlete
Description
Completes the full 4-port calibration. The ports must be calibrated
using the full 1-port calibration in advance. The necessary number
of the thru measurement must be accomplished by the command
SENS:CORR:COLL:THRU:ADD:FULL4:ACQ.
This command calculates the error terms Et and El using the
unknown thru algorithm. Then it saves the Et and El error terms to
the existing calibration getting the full 4-port calibration from the
four 1-port calbrations.
If the full 4-port calibration already existed, updates the Et and El
error terms.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Equivalent Softkeys
Calibration > Calibrate > Thru Addition > Comlete 4-Port Calibration >
Apply