EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sens:corr:coll:meth:trl3; Sens:corr:coll:meth:trl4

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
268
SENS:CORR:COLL:METH:TRL3
SENSe
<Ch>
:CORRection:COLLect:METHod:TRL3
<port1>
,
<port2>
,
<port3>
Description
Selects the ports and sets the 3port TRL calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > 3Port TRL Cal
SENS:CORR:COLL:METH:TRL4
SENSe
<Ch>
:CORRection:COLLect:METHod:TRL4
<port1>
,
<port2>
,
<port3>
,
<port4>
Description
Selects the ports and sets the 4port TRL calibration type for the
calculation of the calibration coefficients on completion of the
calibration executed by the SENS:CORR:COLL:SAVE command.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
<port4> the fourth port number from 1 to 4
Out of Range
If the same port numbers are specified, an error occurs.
Related Commands
SENS:CORR:COLL:SAVE
Equivalent Softkeys
Calibration > Calibrate > 4Port TRL Cal

Table of Contents