EasyManua.ls Logo

Copper Mountain Technologies S2 Series - Sens:corr:trig:free

Copper Mountain Technologies S2 Series
428 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S2VNA and S4VNA SCPI Programming Manual
308
SENS:CORR:TRIG:FREE
SENSe
<Ch>
:CORRection:TRIGger:FREE[:STATe] {OFF|ON|0|1}
SENSe
<Ch>
:CORRection:TRIGger:FREE[:STATe]?
Description
Enables/disables the internal trigger source for calibration. If the
internal trigger source for calibration is enabled then a command
of the calibration standard measurement starts the measurement
immediately. If the internal trigger source for calibration is
disabled then the system trigger source is used (which is set for
regular measurement with the command TRIG:SOUR) to start the
calibration standard measurement.
The system trigger source also enables the averaging trigger
function (TRIG:AVER) and the point trigger function (TRIG:POIN) for
calibration.
Note: When the system trigger source is selected you should avoid
the program trigger source (BUS), otherwise the program deadlock
is possible.
Note: The command does not apply to the electronic calibration,
the power calibration and the receiver calibration. The internal
trigger always used in these cases.
(command/query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
Specifies the trigger source for calibration:
{ON|1} : Internal
{OFF|0} : System
Query Response
{0|1}
Preset Value
1
Equivalent Softkeys
Calibration > Cal Trig Source { Internal | System }

Table of Contents