S2VNA and S4VNA SCPI Programming Manual
214
SENS:CORR:COEF:SAVE
SENSe
<Ch>
:CORRection:COEFficient:SAVE
Description
Enables the written calibration coefficients depending on the
selected calibration type.
On completion of the command the error correction automatically
turns ON.
At the attempt to execute this command before all the needed
calibration coefficients are written, an error occurs and the
command is ignored.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Related Commands
Calibration type selection:
SENS:CORR:COEF:METH:ERES
SENS:CORR:COEF:METH:OPEN
SENS:CORR:COEF:METH:SHOR
SENS:CORR:COEF:METH:THRU
SENS:CORR:COEF:METH:SOLT1
SENS:CORR:COEF:METH:SOLT2
Calibration coefficient writing:
SENS:CORR:COEF
Equivalent Softkeys
None