S2VNA and S4VNA SCPI Programming Manual
292
SENS:CORR:EXT:PORT:LOSS
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:LOSS{[1]|2}
<numeric>
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:LOSS{[1]|2}?
Description
Sets or reads out the values of the loss 1 and loss 2 for the port
extension function.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<numeric> the loss value from 200 to 200
Unit
dB (decibel)
Out of Range
Sets the value of the limit, which is closer to the specified value.
Query Response
<numeric>
Preset Value
0
Equivalent Softkeys
Calibration > Port Extensions > Loss > { Loss1 | Loss2 }