S2VNA and S4VNA SCPI Programming Manual
303
SENS:CORR:REC:COLL:RCH:ACQ
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:COLLect:RCHannel:ACQuire
Description
Executes receiver calibration of the reference receiver of the
specified port <Pt>.The reference receiver calibration uses its own
port <Pt> as the source port.
(no query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Notes
The command starts the measurement of the calibration data
immediately regardless the trigger settings. The command blocks the
execution of the subsequent commands until the completion of the
measurement.
Equivalent Softkeys
Calibration > Receiver Calibration > Take Cal Sweep
SENS:CORR:REC:COLL:TCH:ACQ
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:COLLect:TCHannel:ACQuire
<srcport>
Description
Executes receiver calibration of the test receiver of the specified
port <Pt>.The test receiver calibration uses port number
<srcport> as the source port.
(no query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
<srcport> the number of the source port from 1 to 2 (or 4)
Notes
The command starts the measurement of the calibration data
immediately regardless the trigger settings. The command blocks the
execution of the subsequent commands until the completion of the
measurement.
Equivalent Softkeys
Calibration > Receiver Calibration > Calibrate Both