S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:REC:COLL:RCH:ACQ
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:COLLect:RCHannel:ACQuire
Executes receiver calibration of the reference receiver of the
specified port <Pt>.The reference receiver calibration uses its own
port <Pt> as the source port.
(no query)
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
The command starts the measurement of the calibration data
immediately regardless the trigger settings. The command blocks the
execution of the subsequent commands until the completion of the
measurement.
Calibration > Receiver Calibration > Take Cal Sweep
SENS:CORR:REC:COLL:TCH:ACQ
SENSe
<Ch>
:CORRection:RECeiver
<Pt>
:COLLect:TCHannel:ACQuire
<srcport>
Executes receiver calibration of the test receiver of the specified
port <Pt>.The test receiver calibration uses port number
<srcport> as the source port.
(no query)
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
<srcport> the number of the source port from 1 to 2 (or 4)
The command starts the measurement of the calibration data
immediately regardless the trigger settings. The command blocks the
execution of the subsequent commands until the completion of the
measurement.
Calibration > Receiver Calibration > Calibrate Both