S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:THRU:ADD:FULL3:COMP
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL3:COMPlete
Completes the full 3-port calibration between the ports specified
by the command SENS:CORR:COLL:THRU:ADD:FULL3:PORT. The ports
must be calibrated using the full 1-port calibration in advance. The
necessary number of the thru measurement must be accomplished
by the command SENS:CORR:COLL:THRU:ADD:FULL3:ACQ.
This command calculates the error terms Et and El using the
unknown thru algorithm. Then it saves the Et and El error terms to
the existing calibration getting the full 3-port calibration from the
three 1-port calbrations.
If the full 3-port calibration already existed, updates the Et and El
error terms.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
Calibration > Calibrate > Thru Addition > Comlete 3-Port Calibration >
Apply
SENS:CORR:COLL:THRU:ADD:FULL4:ACQ
SENSe
<Ch>
:CORRection:COLLect:THRU:ADDition:FULL4:ACQuire
<port1>,<port2>
Measures an arbitrary thru between the specified ports. The
measurements are used to complete the 4-port thru addition
function by the command SENS:CORR:COLL:THRU:ADD:FULL4:COMP.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
Calibration > Calibrate > Thru Addition > Comlete 4-Port Calibration >
Thru m-n