SENS:CORR:EXT:PORT:INCL
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:INCLude{[1]|2}[:STATe]
{OFF|ON|0|1}
SENSe
<Ch>
:CORRection:EXTension:PORT
<Pt>
:INCLude{[1]|2}[:STATe]?
Description
Turns ON/OFF the loss compensation of the loss 1 and loss 2 for
the port extension function.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Port Extensions > Loss > { Loss1 | Loss2 }