S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:THRU
SENSe
<Ch>
:CORRection:COLLect[:ACQuire]:THRU
<rcvport>
,
<srcport>
Measures the calibration data of the thru standard between the
receiver port <rcvport> and the source port <srcport>.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<rcvport> the number of the receiver port from 1 to 2 (or 4)
<srcport> the number of the source port from 1 to 2 (or 4)
If the same port numbers are specified, an error occurs.
The command starts the measurement immediately if the trigger source
for calibration set to the "Internal" by the command
SENS:CORR:TRIG:FREE, otherwise waits for the trigger signal. The
command blocks the execution of the subsequent commands until the
completion of the measurement.
Calibration > Calibrate > Response (Thru) > Thru
Calibration > Calibrate > One Path 2–Port Cal > Thru
Calibration > Calibrate > Full 2–Port Cal > Port 1–2 Thru