S2VNA and S4VNA SCPI Programming Manual
358
SOUR:POW:PORT:CORR:COLL:TABL:LOSS
SOURce
<Ch>
:POWer:PORT
<Pt>
:CORRection:COLLect:TABLe:LOSS[:STATe]
{OFF|ON|0|1}
SOURce
<Ch>
:POWer:PORT
<Pt>
:CORRection:COLLect:TABLe:LOSS[:STATe]?
Description
Turns ON/OFF the state of the loss compensation used when the
power calibration is executed by the SOUR:POW:PORT:CORR:COLL
command.
(command/query)
Target
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2} for S2VNA or {[1]|2|3|4} for S4VNA
Parameter
{ON|1} : ON
{OFF|0} : OFF
Query Response
{0|1}
Preset Value
0
Equivalent Softkeys
Calibration > Power Calibration > Loss Compen > Compensation