S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COEF:METH:SOLT4
SENSe
<Ch>
:CORRection:COEFficient:METHod:SOLT4
<port1>
,
<port2>
,
<port3>
,
<port4>
Selects the ports and sets the full 4–port calibration type, when the
written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<port1> the first port number from 1 to 4
<port2> the second port number from 1 to 4
<port3> the third port number from 1 to 4
<port4> the fourth port number from 1 to 4
If the same port numbers are specified, an error occurs.
SENS:CORR:COEF:METH:THRU
SENSe
<Ch>
:CORRection:COEFficient:METHod[:RESPonse]:THRU
<rcvport>
,
<srcport>
Selects the ports and sets the response calibration (Thru) type, when
the written calibration coefficients are made effective by the
SENS:CORR:COEF:SAVE command.
(no query)
Channel <Ch>,
<Ch>={[1]|2|...16}
<rcvport>, the number of the receiver port from 1 to 2 (or 4)
<srcport>, the number of the source port from 1 to 2 (or 4)
If the same port numbers are specified, an error occurs.