301
SENS:CORR:OFFS:COLL:SAVE
SENSe
<Ch>
:CORRection:OFFSet:COLLect:SAVE
Description
Calculates the calibration coefficient for the selected calibration
type (scalar mixer calibration only) from the calibration data
measured with the frequency offset feature is ON.
If this command is executed before all necessary calibration data
for calculating the calibration coefficient is measured, an error
occurs when executed.
(no query)
Target
Channel <Ch>,
<Ch>={[1]|2|...16}
Related Commands
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.METHod.SMIX2
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.LOAD
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.OPEN
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.SHORt
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.THRU
SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.ACQuire.PMETer
Equivalent Softkeys
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration >
Apply