S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:CKIT:ORD:OPEN
SENSe:CORRection:COLLect:CKIT:ORDer:OPEN
<port>
,
<numeric>
SENSe:CORRection:COLLect:CKIT:ORDer:OPEN?
<port>
Sets or reads out the number of the calibration standard of the
open type, used for the measurement of the specified port.
(command/query)
<port> the number of the port from 1 to 2 (or 4)
<numeric> the number of the calibration standard
If the specified standard number is greater than the number of standards
in the kit, an error occurs. If the specified standard number is not the
open standard number, an error occurs.
Calibration > Cal Kit > Specify CLSs > Open Port x (Row)
SENS:CORR:COLL:CKIT:ORD:SEL
SENSe:CORRection:COLLect:CKIT:ORDer:SELect
<numeric>
SENSe:CORRection:COLLect:CKIT:ORDer:SELect?
The subclass used to specify classes of calbration standards by the
commands SENSe<Ch>:CORRection:COLLect:CKIT:ORDer:XXXX.
<numeric> the subclass number from 1 to 8
Calibration > Cal Kit > Specify CLSs > Subclass n (Column)