S2VNA and S4VNA SCPI Programming Manual
SENS:CORR:COLL:CKIT:ORD:THRU
SENSe:CORRection:COLLect:CKIT:ORDer:THRU
<port1>
,
<port2>
,
<numeric>
SENSe:CORRection:COLLect:CKIT:ORDer:THRU?
<port1>
,
<port2>
Sets or reads out the number of the calibration standard of the thru
type, used for the measurement between the <port1> and
<port2> ports.
(command/query)
<port1> the number of the receiver port from 1 to 2 (or 4)
<port2> the number of the source port from 1 to 2 (or 4)
<numeric> the number of the calibration standard
If the specified standard number is greater than the number of standards
in the kit, an error occurs. If the specified standard number is not the thru
standard number, an error occurs.
Calibration > Cal Kit > Specify CLSs > Thru Port x-y (Row)